Water-Detect Feature in the HOBO MX2501 pH Logger is Not Working
{"nid":"7171","uuid":"93908fec-0ac8-4e83-aea4-cc58c6bb4f5e","vid":"8016","langcode":"und","revision_timestamp":"1656615059","status":"1","title":"Water-Detect Feature in the HOBO MX2501 pH Logger is Not Working ","created":"1530554172","changed":"1692131365","promote":"0","sticky":"0","default_langcode":"1","revision_default":"1","revision_translation_affected":"1","path":{"alias":"\/resources\/tech-notes\/water-detect-feature-in-the-hobo-mx2501-ph-logger-not-working","pid":"13851","langcode":"und"},"body":"<p>The HOBO MX2501 pH logger's battery-saving water-detect feature is only guaranteed to work in water with at least 100 \u00b5S\/cm conductivity. The water-detect nodes rely on the water closing the circuit between them. If the water is too fresh, the nodes will not \u201csee\u201d that the logger is in water.<\/p>","applications":["Water Resource Management"],"associated_products":["MX2203","MX2204"],"brand":"HOBO","documentation_hoboconnect":"0","documentation_intemp_app":"0","documentation_is_datasheet":"0","meta_tags":"a:1:{s:6:\"robots\";s:7:\"noindex\";}","product_family_series":"MX Family","resource_topic":["Troubleshooting & Maintenance"],"resource_type":"Tech Note","tech_note_audience":["External"],"tech_note_type":["General Reason"]}
The HOBO MX2501 pH logger's battery-saving water-detect feature is only guaranteed to work in water with at least 100 µS/cm conductivity. The water-detect nodes rely on the water closing the circuit between them. If the water is too fresh, the nodes will not “see” that the logger is in water.