Custom pH Buffers to Calibrate the HOBO MX2501 pH Logger
{"nid":"7151","uuid":"3568e64b-280d-4236-b4f6-12e54897b24e","vid":"7996","langcode":"und","revision_timestamp":"1656606344","status":"1","title":"Custom pH Buffers to Calibrate the HOBO MX2501 pH Logger","created":"1530553301","changed":"1692131342","promote":"0","sticky":"0","default_langcode":"1","revision_default":"1","revision_translation_affected":"1","path":{"alias":"\/resources\/tech-notes\/custom-ph-buffers-to-calibrate-the-hobo-mx2501-ph-logger","pid":"13831","langcode":"und"},"body":"<p>The HOBO MX2501 pH logger calibration procedure is only designed to work with National Bureau of Standards (NBS) pH buffers 4.01, 7.00, and 10.00. However, users can check the calibration of the logger in any buffer that is at a similar temperature to the logger\u2019s calibration temperature and that falls within the operating pH range of the logger (pH 2-12).<\/p>","applications":["Water Resource Management"],"associated_products":["MX2501"],"brand":"HOBO","documentation_hoboconnect":"0","documentation_intemp_app":"0","documentation_is_datasheet":"0","meta_tags":"a:1:{s:6:\"robots\";s:7:\"noindex\";}","product_family_series":"MX Family","resource_topic":["Getting Started","Troubleshooting & Maintenance"],"resource_type":"Tech Note","tech_note_audience":["External"],"tech_note_type":["Measurement and accuracy"]}
The HOBO MX2501 pH logger calibration procedure is only designed to work with National Bureau of Standards (NBS) pH buffers 4.01, 7.00, and 10.00. However, users can check the calibration of the logger in any buffer that is at a similar temperature to the logger’s calibration temperature and that falls within the operating pH range of the logger (pH 2-12).